Medical Device Symbols and Labelling

IEC 60601-1 marking requirements

ISO 15223-1 and IEC 60601-1 Marking

A patient monitor passes its electrical safety measurements, but the laboratory identifies an incorrect Type BF marking and outdated voltage ratings on its equipment label. The manufacturer must correct these issues before compliance can be demonstrated.

For medical equipment manufacturers in India, IEC 60601-1 marking requirements are an important part of electrical safety evaluation, not simply a packaging consideration.

ISO 15223-1 standardizes symbols used to communicate medical device information. IEC 60601-1 Clause 7 establishes safety-related equipment marking, identification, and accompanying-document requirements. Both can apply to the same device alongside Indian regulatory labelling requirements.

ISO 15223-1 vs IEC 60601-1: What Is the Difference?

ISO 15223-1:2021 covers medical device symbols for identification, traceability, handling, and other manufacturer-supplied information. Amendment 1:2025 updates the authorized representative definition and associated symbol.

The 2021 edition includes the Medical Device (MD) symbol and Unique Device Identifier (UDI) symbol.

For example, a reusable patient monitor may carry manufacturer, serial number, MD, and applicable UDI information. Its separately supplied single-use SpO2 sensor may require additional single-use or sterility-related markings, depending on its intended use and supplied condition.

IEC 60601-1:2005 + AMD1:2012 + AMD2:2020 (Edition 3.2), however, addresses electrical equipment safety markings.

Its Clause 7 requirements apply to equipment identification, supply ratings, applied parts, safety information, and accompanying documents.

IEC 60601-1 Marking Requirements Under Clause 7

Clause 7 identifies markings necessary for safe equipment identification and operation.

Depending on the equipment, manufacturers must address electrical ratings, protective earth terminals, Class II identification, applied-part markings, operating limitations, and enclosure protection ratings.

Where Do IEC 60601-1 Symbols Come From?

IEC 60601-1 Annex D collects symbols referenced by the standard, including symbols originating from IEC 60417, ISO 7000, and ISO 7010.

Marking

Reference

Purpose

Class II

IEC 60417-5172

Double-insulated equipment

Type BF

IEC 60417-5333

BF applied-part classification

Type CF

IEC 60417-5335

CF applied-part classification

Protective earth

IEC 60417-5019

Protective earth terminal

Defibrillation-proof BF

IEC 60417-5334

Defibrillation-proof BF applied part

Follow instructions

ISO 7010-M002

Mandatory instruction sign

Manufacturers must use the appropriate registered graphical symbols rather than substituting similar-looking artwork.

The IEC 60601-1 compliance guide explains the broader safety framework.

Consult Instructions vs Follow Instructions: A Common Marking Mistake

The ISO 15223-1 Consult instructions for use symbol and ISO 7010-M002 Follow instructions for use sign communicate different requirements.

The consult symbol, associated with ISO 7000-1641, directs users to the instructions for information.

ISO 7010-M002 is a mandatory safety sign, represented by a white reading figure inside a blue circle.

Where IEC 60601-1 requires following the instructions as a safety measure, substituting the ordinary consult symbol may create a marking nonconformity.

Manufacturers should verify whether their risk controls require a mandatory safety sign rather than a general reference to documentation.

Applied-Part Marking and Placement Requirements

IEC 60601-1 distinguishes Type B, BF, and CF applied parts based on patient protection characteristics.

The appropriate symbol must identify the relevant applied part or its connection point according to Clause 7.2.10.

For example, a patient monitor with a Type BF SpO2 connection should display the applicable classification at the appropriate location.

If a connector supports applied parts with different classifications, or electrical isolation depends on the applied part itself, marking placement must reflect that arrangement.

Defibrillation-proof applied parts require their corresponding symbols. Additional warnings may apply when defibrillation protection depends on a specified patient cable.

Manufacturers can review Type B, BF, and CF applied-part classifications for further guidance.

Label Durability and Legibility Testing

IEC 60601-1 Clause 7.1.2 addresses marking legibility under specified viewing conditions.

Clause 7.1.3 addresses durability. Applicable markings undergo rubbing tests using distilled water, methylated spirit, and isopropyl alcohol, each for a specified period.

The markings must remain legible, and adhesive labels must not become loose or curl at their edges.

A label that fades or detaches during testing can create a nonconformity requiring correction and re-verification.

Other Marking Failures Manufacturers Should Check

An equipment power supply change can invalidate previously approved label artwork if the rated input information no longer matches the finished design.

Likewise, equipment claiming ingress protection may require an appropriate IP code, such as IPX1, consistent with its evaluated protection level.

Class II equipment must carry the correct classification marking, while Class I equipment requires appropriate identification of protective earth connections.

Manufacturers should compare final label artwork against the electrical design, applied-part classification, and instructions for use.

The IEC 60601-1 pre-test documentation checklist provides useful preparation guidance.

Medical Device Labelling Requirements in India

Rule 44 of India’s Medical Devices Rules, 2017 establishes applicable labelling particulars, including device identification, manufacturer information, manufacturing details, and licence particulars.

Recognized BIS or ISO symbols may replace certain text where permitted, provided users can understand the information safely.

ISO 15223-1 symbols therefore complement Indian regulatory requirements rather than replacing them.

IEC 60601-1 Marking Assessment During Medical Device Testing

Astute Labs provides medical device testing services against IEC 60601-1 and applicable IS 13450 requirements.

Marking and accompanying-document checks form part of the applicable IEC 60601-1 safety evaluation. Reviewing label artwork alongside equipment construction helps manufacturers identify potential nonconformities before final submission.

Getting Equipment Markings Right Before Testing

Accurate medical device labelling requires consistency between standardized symbols, electrical safety markings, technical documentation, and applicable regulatory requirements.

Correcting marking errors before formal evaluation can reduce avoidable delays and repeated verification.

For assistance with IEC 60601-1 electrical safety and marking requirements, Contact us.

Frequently asked questions

01. What does IEC 60601-1 Clause 7 cover?
Clause 7 addresses identification, equipment markings, safety symbols, and accompanying documents for medical electrical equipment.
No. They originate from IEC 60417 and are referenced by IEC 60601-1 for applied-part classification.
Yes. M002 is a mandatory instruction sign, while the consult symbol directs users to the instructions for information.
Yes. Incorrect ratings, missing symbols, poor marking placement, or failed durability tests can result in compliance nonconformities.  
Not independently. Manufacturers must also satisfy applicable requirements under India’s Medical Devices Rules, 2017.

About Author

Yash Chawlani is your go-to digital marketing specialist and founder of Merlin Marketing, a performance-driven marketing agency. With over 7 years of experience, Yash has worked with some big names like Elementor, G2, and Snov, just to name a few, to boost their online presence. When he's not diving into the latest marketing trends, you'll either find him at the gym or on the football field.