Harmonics and Flicker Testing for Medical Devices

harmonics and flicker testing

IEC 61000-3-2 and 61000-3-3 Explained

A dialysis machine passes conducted emissions testing but exceeds the permitted fifth-harmonic current limit. Another medical device experiences excessive voltage changes when its compressor starts, causing a flicker test failure.

Both issues originate from equipment interacting with the electrical supply, but they involve different electrical characteristics.

For medical device manufacturers in India, harmonics and flicker testing helps establish whether equipment meets applicable low-frequency EMC emission requirements.

IEC 61000-3-2 limits harmonic currents drawn from public electricity supplies, while IEC 61000-3-3 limits voltage changes and flicker caused by changing electrical loads. Applicability depends on the device’s rated power, input current, connection arrangement and relevant exemptions.

What Is Harmonic Current Testing Under IEC 61000-3-2?

Harmonics occur when nonlinear electrical circuits draw distorted current from the AC supply.

Switching power supplies, rectifiers and power conversion circuits can produce current components at multiples of the fundamental frequency.

For India’s nominal 50 Hz mains frequency, the third harmonic occurs at 150 Hz and the fifth at 250 Hz.

IEC 61000-3-2:2018 + AMD1:2020 + AMD2:2024 (Edition 5.2) establishes harmonic current limits for applicable equipment rated up to 16 A per phase.

IEC 60601-1-2 references harmonic emission requirements for medical electrical equipment connected to the public mains network.

IEC 61000-3-2 Class A Harmonic Limits

For applicable Class A equipment, harmonic limits are specified as absolute current values in amperes.

Harmonic order

Maximum current

3rd

2.30 A

5th

1.14 A

7th

0.77 A

9th

0.40 A

11th

0.33 A

These are representative Class A limits. The complete standard includes additional harmonics, classifications and test conditions.

A higher-power medical device using an inadequately designed switching supply can draw narrow current pulses that exceed applicable limits.

Manufacturers can reduce harmonic distortion through passive power factor correction (PFC), such as input inductors, or active PFC circuits that shape input current more effectively.

The appropriate solution depends on equipment power, efficiency, size and design constraints.

When Do the 75 W and 1 kW Harmonic Exemptions Apply?

Not every mains-powered medical device has harmonic current limits.

Under IEC 61000-3-2, equipment with rated power of 75 W or less generally has no specified harmonic current limits, except for applicable lighting equipment.

This distinction matters for low-power monitoring and diagnostic equipment.

The standard also specifies no Class A harmonic limits for qualifying professional equipment with total rated power above 1 kW.

However, manufacturers must verify that the equipment satisfies the standard’s professional-equipment definition and applicable classification requirements.

These exemptions concern harmonic limits, not a blanket exemption from IEC 60601-1-2 EMC requirements.

What Is Flicker Testing Under IEC 61000-3-3?

Flicker testing evaluates voltage fluctuations caused by changing electrical loads.

For example, a medical compressor can draw a high starting current, causing a temporary voltage drop. Repeated heater switching or motor operation may also create voltage changes affecting other equipment connected to the supply.

IEC 61000-3-3:2013 + AMD1:2017 + AMD2:2021 (Edition 3.2) applies to qualifying equipment rated up to 16 A per phase, connected to public low-voltage networks and not subject to conditional connection.

Its specified 220 V to 250 V line-to-neutral, 50 Hz supply range is relevant to India’s nominal 230 V, 50 Hz electrical system.

Important IEC 61000-3-3 Flicker Parameters

Parameter

Measurement

Common limit

Pst

Short-term flicker severity, typically 10 minutes

≤ 1.0

Plt

Long-term flicker severity, typically 2 hours

≤ 0.65

dc

Steady-state relative voltage change

≤ 3.3%

dmax

Maximum relative voltage change

Generally ≤ 4%

The Plt requirement is particularly relevant to equipment normally operating for more than 30 minutes. Additional conditions and exceptions apply to specific switching characteristics and voltage-change limits.

Motor and compressor inrush currents can increase dmax, while repeated heater cycling can influence flicker severity.

Soft-start circuits, staged switching and controlled power delivery can reduce these disturbances.

Can Medical Equipment Be Exempt from Flicker Testing?

IEC 61000-3-3 permits testing to be omitted when equipment is unlikely to produce significant voltage fluctuations or flicker.

This determination should be supported by examination of the equipment design, electrical specifications and, where necessary, a short functional assessment.

For example, a constant-load medical instrument without significant switching or inrush events may qualify for a justified assessment without full flicker testing.

Manufacturers should document the technical basis rather than assuming that low-power equipment is automatically exempt.

Harmonics vs Flicker Testing: Key Differences

Parameter

Harmonics

Flicker

Standard

IEC 61000-3-2

IEC 61000-3-3

Main measurement

Harmonic current components

Voltage fluctuations

Typical cause

Nonlinear power conversion

Changing electrical loads

Common failure

Excessive third or fifth harmonic

Excessive Pst or dmax

Typical improvement

Passive or active PFC

Soft start and controlled switching

Both are emission assessments. They differ from immunity testing, which evaluates equipment resistance to external disturbances.

Planning Harmonics and Flicker Compliance Under IEC 60601-1-2

IEC 60601-1-2:2014 + AMD1:2020 (Edition 4.1) addresses electromagnetic compatibility for medical electrical equipment.

Its Clause 7.2 connects harmonic and flicker requirements to applicable public mains connections, while Table 2 distinguishes professional healthcare and home healthcare environments.

Manufacturers should identify rated power, mains voltage, current, load profiles and operating configurations before determining test applicability.

A review of power supply architecture during development can also help identify harmonic risks before component selection becomes difficult to change.

The IEC 60601-1-2 medical EMC testing guide provides additional information on the overall evaluation requirements.

Medical Equipment EMC Planning at Astute Labs

Astute Labs provides EMI/EMC testing services for medical and electronic equipment, including pre-compliance evaluations.

For manufacturers, early identification of applicable emissions requirements helps avoid unnecessary testing and reduces the risk of redesigning power circuitry after development.

The EMC pre-compliance testing guide explains the value of identifying emissions problems before formal evaluation.

Choosing the Right EMC Emission Test Plan

Harmonics and flicker testing address different effects of medical equipment on the shared electrical supply.

Understanding the 75 W rule, professional-equipment provisions, voltage fluctuation parameters and public mains applicability helps manufacturers establish an appropriate test plan.

For assistance with medical device EMC compliance planning and testing requirements, Contact us.

Frequently asked questions

01. What is the difference between harmonics and flicker testing?
Harmonic testing measures distorted current drawn from the mains supply. Flicker testing evaluates voltage fluctuations caused by changing equipment loads.
Equipment rated at 75 W or less generally has no harmonic current limits under IEC 61000-3-2, except for applicable lighting equipment. Other EMC requirements can still apply.
Qualifying professional equipment above 1 kW may have no Class A harmonic limits. Manufacturers must verify the equipment classification and applicable conditions.
Pst measures short-term flicker severity over approximately 10 minutes, while Plt evaluates long-term flicker severity over approximately two hours.  
Not always. IEC 61000-3-3 permits testing to be omitted when a justified technical assessment shows that significant voltage fluctuations or flicker are unlikely.

About Author

Yash Chawlani is your go-to digital marketing specialist and founder of Merlin Marketing, a performance-driven marketing agency. With over 7 years of experience, Yash has worked with some big names like Elementor, G2, and Snov, just to name a few, to boost their online presence. When he's not diving into the latest marketing trends, you'll either find him at the gym or on the football field.

Blogs

From the knowledge hub

Oxygen Concentrator Safety Testing: ISO 80601-2-69
View Blog
harmonics and flicker testing
View Blog
ESD EFT surge immunity testing
View Blog
Scroll to Top